This information is only relevant to developers engaded in developing and maintaining the API implementation. It also may be relevant to those creating system maintenance and security/architecture documentation.
Perhaps the most notable implication of the implementation is that some of the configuration information that appears (when viewed through the API) as being stored in the Unit is actualy stored in the chairside PC.
Having said that, here goes.
The current production Dental Units do not have the horse power neither in hardware or software to handle TCP/IP directly. Therefore a lot of the functionality of the API is implemented by code running in the chair side PC that connects to the Unit via Planmeca proriaritary serial protocol called Planet. This protocol is not available for application developers and is not supported.
Some API functionality is also implemented in the client side API Java code, notably inside the DentalUnit class.
Below the high level PMUAPI internal achitecture is illustrated.